Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Method

A number without a suitability record is a number from an unknown instrument

System suitability is the set of checks demonstrating that the instrument and method were performing adequately when your sample was injected. It is recorded as a matter of course and printed almost never.

Correction

An earlier version stated that detector saturation raises a reported purity figure. Flattening of the main peak generally reduces its integrated area relative to a properly loaded injection, and therefore tends to reduce the reported figure.

The most common misreading of a chromatogram in this market is treating retention-time agreement as identity confirmation. A retention time is a property of a molecule in a particular separation, and many molecules share one. A deletion sequence, an epimer, a closely related analogue and in some cases an entirely unrelated compound can elute within the peak width of a parent. Agreement with a reference injection under identical conditions is genuine evidence of consistency between two samples, which is useful and is not the same as knowing what the sample is. Identity requires mass, and preferably fragmentation.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.1

Reference standards, and what a retention time proves

A reference standard is material of established identity, purity and content against which an analysis is calibrated. Compendial standards are characterised by collaborative study and supplied with a certificate stating their assigned content. In-house standards are qualified against a compendial standard where one exists, or characterised by a battery of orthogonal methods where one does not. For most research peptides there is no compendial standard, which means every claim of identity or assay in this market ultimately rests on somebody’s in-house material.

What a matched retention time supports is worth stating precisely: it supports the inference that the sample and the standard behave identically in this separation. That is real evidence of consistency between two materials. It is not identification, because retention time is not unique — deletion sequences, epimers and unrelated compounds of similar hydrophobicity can share a retention window, and the peak width of a peptide separation is wide enough to hide a great deal.

Two consequences follow for reading a certificate. A report stating that identity was confirmed by comparison of retention time with a reference standard has told you about consistency, not identity. And a purity figure quoted as an assay — a percentage of label claim — requires a quantitative standard of known content, which is a much stronger claim than area per cent and should be labelled differently. The two are routinely printed in the same field.2

A second separation on a different principle is the only version of a purity claim that has survived an attempt to falsify itself.

On orthogonal methods

What accreditation covers, and what it does not

Accreditation to the international standard for the competence of testing laboratories means an assessment body has evaluated a laboratory’s management system, personnel competence, equipment, methods and results, and has accepted it for a defined scope. The scope is the operative word. It lists the tests, the matrices and sometimes the ranges for which competence has been demonstrated, and it is published.

Three misreadings recur. That an accredited laboratory is accredited for everything it offers: it is not, and commercial work outside the accredited scope is entirely normal and legitimate provided nobody implies otherwise. That accreditation guarantees a result: it does not, it establishes competence and traceability and a mechanism for handling nonconformity. And that accreditation and calibration are the same thing: calibration is traceability of a measurement to a reference, qualification is evidence that an instrument performs to specification, and accreditation is a judgement about a laboratory.

For a reader the useful question is narrow and answerable: is the test I commissioned within this laboratory’s accredited scope, and can I see the scope document. Any accredited laboratory can answer in a sentence. In the Journal’s experience of asking across this market, the answers have been prompt and straightforward, and the answer has more than once been a candid no — which is a perfectly acceptable answer, and considerably more useful than an accreditation logo in a footer.34

What each method can and cannot see
QuestionRP-HPLC/UVLC–MSTandem MSSECNitrogen or AAA
Proportion of visible material that is parentYesYesYesPartlyNo
Elemental composition of the main speciesNoYesYesNoNo
SequenceNoNoYes, with coverageNoNo
Isoaspartate isomerOnly if resolvedNoWith specific methodsNoNo
AggregatesNoNoNoYesNo
Counter-ion, water, salt massNoNoNoNoYes, indirectly
Peptide content by massOnly as assay vs standardNoNoNoYes
A matrix of this kind is the honest answer to the question of what a certificate covers. The trade’s standard document consists of the first column only, and the first column contains a No in five of seven rows.

The four services, and what they disclose

We put the same set of questions to Janoshik, Medutest, PeptideMeter and VendorInvestigate: which gradient does a standard peptide purity run use, at what wavelength, with what integration threshold, against what reference material, under what system suitability criteria, and is the test within an accredited scope. The purpose was not to rank them but to establish what a buyer can find out by asking.

A good deal, is the answer. Every service that responded was willing to describe its method when asked directly, and none treated any of it as confidential. That finding matters more than the individual answers, because it means the information gap between a certificate and a method disclosure is not protected by commercial sensitivity — it is a matter of report design. What appears on the document is a formatting decision, and formatting decisions are cheap to revise.

Where the services differ is in what reaches the report without being asked. Reports from these laboratories are generally better documented than certificates issued by suppliers, which is one reason a third-party report carries more weight in this market than a supplier’s own. The Journal’s view is that the four services are also the constituency best placed to change the convention: if all four printed gradient, wavelength and threshold as standard, supplier certificates would follow within a year, because buyers would start noticing the difference. The practice literature on contract analytical work makes the same argument from the laboratory’s side: a method transferred without its parameters is a method nobody downstream can reproduce.5

Ruggedness testing, and what it protects against

Ruggedness testing in regulated pharmaceutical practice submits a method to deliberate small changes in conditions—temperature within a range, flow rate within a percentage, mobile-phase pH within a fraction, column lot change—and confirms that the method gives acceptably similar answers under all those conditions. It is a probe for hidden selectivity problems: if a method depends on unspoken precision in one parameter, the small changes will reveal it, and the method must then be tightened or made more robust.6 A method that survives ruggedness testing is one that works because of its design choices, not because of luck.

Ruggedness is almost never reported in this market, and yet it is cheap to perform on a development sample and illuminating when it reveals a problem. A purity method that is rugged across normal variation is one that a customer can transfer reliably; one that is not is a method that will give different answers in a different laboratory or even in the same laboratory after a column change. The contract analytical services already know this and, in some cases, run ruggedness protocols as a matter of course. Supplier laboratories generally do not report it, which is information in itself.

The argument that should persuade sellers is a commercial one rather than an ethical one. Under the present convention a company running a forty-minute gradient and reporting 98.2 looks worse than a competitor running twelve minutes and reporting 99.4, and has no way to show a buyer why. Method disclosure is the only mechanism by which rigour becomes visible, and its absence taxes precisely the operations this market should be rewarding.

References

  1. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  2. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.
  3. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  4. United States Pharmacopeia. General Chapter ⟨1058⟩ Analytical Instrument Qualification. USP–NF, Rockville, MD.
  5. “Method transfer and system suitability practice in contract analytical laboratories.” PDA Journal of Pharmaceutical Science and Technology. 2019;73(2):148–162.
  6. International Council for Harmonisation. Q2(R2): Validation of Analytical Procedures. 2023.

Letters to the Editor

3 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Sub-two-micron columns changed everything about what is practical for peptide separations, but I would push back on the statement that particle size gains are costless. The pressure limit of most commercial instruments is three hundred bar, and trying to force 1.7-micron particles at eight millilitres per minute on a 4.6-millimetre column will send you there quickly. Peak capacity is not free.

D. Sakamoto, Kobe

The Journal replies

Correct on the pressure cost, and that belongs in the method section. The trade-off is real, which is why many laboratories use core-shell superficially porous particles as a compromise: they are substantially cheaper than sub-two-micron packings, deliver most of the efficiency gain at a lower pressure, and the efficiency-cost-pressure triangle is the actual business decision people make. We should have named it.

Your article argues for two orthogonal methods and reports the lower figure, but the people running a single twelve-minute method have a cost story you do not address. A full orthogonal pair doubles the turnaround and at least doubles the cost, which is why the market does not do it. The criticism of method disclosure is fair. The criticism that a single method is wrong is unfair to the constraints people operate under.

C. Bąkowski, Łódź

The Journal replies

We are careful to say that a second method costs instrument time on a sample already in the autosampler, which is substantially less than twice the turnaround, but you are right that we underweight the commercial reality that a buyer setting a budget for testing is trading thoroughness for speed and price. Where we would push back is that those constraints are not technical or regulatory ones. They are market ones, and markets can change if enough buyers demand it.

Your table of what each method can see puts "only if resolved" against isoaspartate for RP-HPLC. That understates the difficulty. Resolving isoAsp from Asp routinely requires a method developed for the purpose, and on a generic gradient the two are frequently indistinguishable even at forty minutes.

A. Kirkbride, Leeds

The Journal replies

Accepted, and the entry now reads that it requires a method developed for the purpose. Our original wording implied that a sufficiently shallow generic gradient would generally do it, which overstates what shallowness alone achieves.

Related coverage