Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Purity

Mass spectrometry answers a different question, properly

Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.

The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.1

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.2

Reversed-phase chromatography answers its own question superbly. The market asks it five questions and prints one answer.

Orla McCaffrey, Staff Writer, Analytics

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

Where two figures on the same lot disagree, the useful first question is not which laboratory is right but which gradient was shorter. In every disagreement this desk has been able to decompose, the faster method returned the higher number, and the difference was reconstructible from the methods alone.

Detection wavelength and what responds to it
WavelengthPrincipal absorberSees fragments without aromaticsBaseline noiseTypical use
214 nmAmide bondYesHigherPeptide purity and related substances
220 nmAmide bondYesModeratePeptide purity, quieter baseline
254 nmAromatic systemsNoLowSmall-molecule work, legacy detectors
280 nmTrp, Tyr, Phe side chainsNoLowTracking an aromatic residue; not a purity method
Diode array, 200–400 nmAll of the aboveYesMethod-dependentPeak purity assessment, spectral homogeneity
A purity figure generated at 280 nm is not comparable with one generated at 214 nm, and the difference is not a matter of a percentage point. Certificates reading at 280 nm circulate in this market; readers have sent us several.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

The full spectrum, recorded and discarded

A diode-array detector records ultraviolet absorbance across a wavelength range—commonly 200 to 400 nanometres—at every point in the chromatogram. From that data, two things of immediate value can be extracted. First, the spectrum can be examined at each peak to confirm spectral homogeneity—a pure peak has a spectrum that does not drift across its width, while a co-eluted pair shows a spectrum that changes. Second, the full spectrum can be plotted to reveal absorbers that were not visible at the single monitoring wavelength.3 A 254-nanometre impurity in a sample monitored at 280 nanometres is invisible in the 214-nanometre chromatogram and is therefore invisible in the purity report, until a diode-array scan reveals it.

Almost no research-peptide certificate prints this information. The data exists—it is generated automatically—and its absence from the report is a formatting decision, not a technical limitation. The usefulness of seeing the full spectrum would be apparent on the first certificate where it resolves an otherwise inexplicable discrepancy between the purity and the mass-balance calculation. It is one of the few changes to the standard format that costs nothing and adds information the reader genuinely needs.

The economics deserve restating, because they explain the convention rather than excusing it. A generic gradient occupies an instrument for under an hour and returns a number the same week. Every one of the determinations that would qualify that number costs several times as much and takes several times as long.

Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.

References

  1. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  2. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.
  3. European Pharmacopoeia. Chapter 2.2.29 — Liquid Chromatography. Council of Europe, Strasbourg.

Letters to the Editor

4 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Two methods are only orthogonal if the mechanisms are independent, and two reversed-phase columns from different vendors are not. We spent a year believing we had orthogonality because the selectivities differed a little, and we had nothing of the kind. The test is the mechanism, not the part number.

M. Quintero, San Juan

Your rule about writing down the lower of two orthogonal figures is sound and I would add the corollary. If the two agree within the repeatability of the method, you have learned something real about the material. If they disagree by three points, you have learned something real about at least one of the methods, and that is worth knowing too.

R. Malinowska, Białystok

The Journal replies

Both halves of that are right, and the second is the half readers tend to discard. A disagreement between two honest methods is information, not noise.

Our practice is to require the chromatogram alongside the number, and the striking thing is how often the request is met and how rarely anyone makes it. Two years of asking and we have been refused twice. The data is sitting on a laboratory server that nobody is querying.

E. Vasquez-Rueda, Cali

A modest defence of the percentage. It is the only figure in this trade that is generated by an instrument rather than asserted by a seller, and for all its limits that is not nothing. I would rather argue about what a number means than have no number to argue about.

L. Silveira, Belo Horizonte

The Journal replies

We agree, and the piece says so in the closing paragraph. The complaint is not that the figure exists but that it circulates stripped of the four values that make it comparable with the next one.

Related coverage