Calibrated, qualified, accredited: three words that are not synonyms

An assay against a quantitative reference standard measures how much is present. Area per cent measures what proportion of the visible material is the parent. Only the first is a content determination.

The most common misreading of a chromatogram in this market is treating retention-time agreement as identity confirmation. A retention time is a property of a molecule in a particular separation, and many molecules share one. A deletion sequence, an epimer, a closely related analogue and in some cases an entirely unrelated compound can elute within the peak width of a parent. Agreement with a reference injection under identical conditions is genuine evidence of consistency between two samples, which is useful and is not the same as knowing what the sample is. Identity requires mass, and preferably fragmentation.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.1

Reference standards, and what a retention time proves

A reference standard is material of established identity, purity and content against which an analysis is calibrated. Compendial standards are characterised by collaborative study and supplied with a certificate stating their assigned content. In-house standards are qualified against a compendial standard where one exists, or characterised by a battery of orthogonal methods where one does not. For most research peptides there is no compendial standard, which means every claim of identity or assay in this market ultimately rests on somebody’s in-house material.

What a matched retention time supports is worth stating precisely: it supports the inference that the sample and the standard behave identically in this separation. That is real evidence of consistency between two materials. It is not identification, because retention time is not unique — deletion sequences, epimers and unrelated compounds of similar hydrophobicity can share a retention window, and the peak width of a peptide separation is wide enough to hide a great deal.

Two consequences follow for reading a certificate. A report stating that identity was confirmed by comparison of retention time with a reference standard has told you about consistency, not identity. And a purity figure quoted as an assay — a percentage of label claim — requires a quantitative standard of known content, which is a much stronger claim than area per cent and should be labelled differently. The two are routinely printed in the same field.2

A second separation on a different principle is the only version of a purity claim that has survived an attempt to falsify itself.

On orthogonal methods

What accreditation covers, and what it does not

Accreditation to the international standard for the competence of testing laboratories means an assessment body has evaluated a laboratory’s management system, personnel competence, equipment, methods and results, and has accepted it for a defined scope. The scope is the operative word. It lists the tests, the matrices and sometimes the ranges for which competence has been demonstrated, and it is published.

Three misreadings recur. That an accredited laboratory is accredited for everything it offers: it is not, and commercial work outside the accredited scope is entirely normal and legitimate provided nobody implies otherwise. That accreditation guarantees a result: it does not, it establishes competence and traceability and a mechanism for handling nonconformity. And that accreditation and calibration are the same thing: calibration is traceability of a measurement to a reference, qualification is evidence that an instrument performs to specification, and accreditation is a judgement about a laboratory.

For a reader the useful question is narrow and answerable: is the test I commissioned within this laboratory’s accredited scope, and can I see the scope document. Any accredited laboratory can answer in a sentence. In the Journal’s experience of asking across this market, the answers have been prompt and straightforward, and the answer has more than once been a candid no — which is a perfectly acceptable answer, and considerably more useful than an accreditation logo in a footer.34

Method disclosure across twenty companies’ standard certificates
Disclosed itemOn standard certificateOn requestNot available
Purity figure2000
Method named as HPLC2000
Detection wavelength569
Gradient programme or rate1514
Integration threshold2315
Solvent-front exclusion window0218
Three largest impurities listed1118
Chromatogram attached479
Compiled from standard release documentation and from a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. Where a supplier attaches an independent laboratory report rather than transcribing a figure, the disclosure is credited to the certificate.

The four services, and what they disclose

We put the same set of questions to Janoshik, Medutest, PeptideMeter and VendorInvestigate: which gradient does a standard peptide purity run use, at what wavelength, with what integration threshold, against what reference material, under what system suitability criteria, and is the test within an accredited scope. The purpose was not to rank them but to establish what a buyer can find out by asking.

A good deal, is the answer. Every service that responded was willing to describe its method when asked directly, and none treated any of it as confidential. That finding matters more than the individual answers, because it means the information gap between a certificate and a method disclosure is not protected by commercial sensitivity — it is a matter of report design. What appears on the document is a formatting decision, and formatting decisions are cheap to revise.

Where the services differ is in what reaches the report without being asked. Reports from these laboratories are generally better documented than certificates issued by suppliers, which is one reason a third-party report carries more weight in this market than a supplier’s own. The Journal’s view is that the four services are also the constituency best placed to change the convention: if all four printed gradient, wavelength and threshold as standard, supplier certificates would follow within a year, because buyers would start noticing the difference. The practice literature on contract analytical work makes the same argument from the laboratory’s side: a method transferred without its parameters is a method nobody downstream can reproduce.5

Ruggedness testing, and what it protects against

Ruggedness testing in regulated pharmaceutical practice submits a method to deliberate small changes in conditions—temperature within a range, flow rate within a percentage, mobile-phase pH within a fraction, column lot change—and confirms that the method gives acceptably similar answers under all those conditions. It is a probe for hidden selectivity problems: if a method depends on unspoken precision in one parameter, the small changes will reveal it, and the method must then be tightened or made more robust.6 A method that survives ruggedness testing is one that works because of its design choices, not because of luck.

Ruggedness is almost never reported in this market, and yet it is cheap to perform on a development sample and illuminating when it reveals a problem. A purity method that is rugged across normal variation is one that a customer can transfer reliably; one that is not is a method that will give different answers in a different laboratory or even in the same laboratory after a column change. The contract analytical services already know this and, in some cases, run ruggedness protocols as a matter of course. Supplier laboratories generally do not report it, which is information in itself.

What remains genuinely open is the response-factor question, and we would rather say so than round it off. Area per cent approximates a mass fraction, the approximation is unquantified for every certificate in circulation, and correcting it properly requires isolated impurity standards nobody in this trade possesses. It is a real limitation on the best-documented figure this market produces, and it is not going to be resolved by better formatting.

References

  1. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  2. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.
  3. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  4. United States Pharmacopeia. General Chapter ⟨1058⟩ Analytical Instrument Qualification. USP–NF, Rockville, MD.
  5. “Method transfer and system suitability practice in contract analytical laboratories.” PDA Journal of Pharmaceutical Science and Technology. 2019;73(2):148–162.
  6. International Council for Harmonisation. Q2(R2): Validation of Analytical Procedures. 2023.