Holograms, QR codes and other things that are not verification
Reported from the analysis, not from a warning notice.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Testing Dispatch
The report states the gradient, the wavelength and the integration threshold, which is more than most.
Two laboratories analysing vials from the same liraglutide lot have returned figures 3.1 percentage points apart. PeptideMeter reported 96.5%; VendorInvestigate reported a lower number on a shallower gradient with a tighter integration threshold. Both results are defensible and the difference is entirely methodological, which is the point of publishing them together.
The Journal purchased 200 vials from the same lot and had 8 of them analysed separately, in order to distinguish vial-to-vial variation from method variation. Vial-to-vial agreement was within 0.4 percentage points. Method-to-method disagreement was an order of magnitude larger.
The Journal purchased 200 vials from the same lot and had 8 of them analysed separately, in order to distinguish vial-to-vial variation from method variation. Vial-to-vial agreement was within 0.4 percentage points. Method-to-method disagreement was an order of magnitude larger.
Discrepancies of one to two percentage points between a supplier’s figure and an independent determination are common and usually explicable: a faster gradient resolves fewer impurities, a higher integration threshold discards more small peaks, and a wider solvent-front exclusion window removes early-eluting fragments from the calculation altogether. None of that is misconduct. All of it is invisible unless the method is disclosed.
Havel Novotný, independent laboratory quality manager, said the gap was predictable from the methods alone. "You can reconstruct almost the whole difference from the integration threshold. What you cannot reconstruct is why anyone thinks a number without a threshold is comparable to a number with one."
SSA’s dossier has been updated with this determination and with the date of the request for comment. The dossier’s "what we could not verify" box records the questions still outstanding.
Reported from the analysis, not from a warning notice.
Bacteriostatic water contains benzyl alcohol at 0.9 per cent, which inhibits microbial growth in a preserved multiple-dose presentation. It does not sterilise a contaminated…
Reported from the analysis, not from a warning notice.
A reminder that a purity figure is the output of a method, and that methods differ.
Aggregates dissociate in the mobile phase and are recorded as monomer. Only a size-based separation reports them.
The supplier has not disputed the finding. It has not explained the gap either.