The qualification chain nobody in this trade can show you
Accreditation to the international competence standard covers the scope a laboratory has been assessed for, which is not necessarily the test you commissioned.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Testing Dispatch
The supplier has not disputed the finding. It has not explained the gap either.
A peptide content determination commissioned by the Journal on a 30 mg mazdutide vial from ERP has returned 97% of the labelled mass. Chromatographic purity on the same vial was above 97.8%. The two numbers measure different things and only one of them was on the certificate.
Reported figures across the set were: highest 97.8%, lowest 93.6%, median close to the midpoint of the two. Every report in the set stated a detection wavelength, and 21 of them stated an integration threshold. The Journal has asked the laboratory to publish thresholds on all reports as a matter of routine.
Content by elemental nitrogen determination returned 97% of the labelled nominal mass, which on a 30 mg vial corresponds to roughly 8.68 mg of peptide rather than the labelled figure. The remainder is accounted for by counter-ion, residual water and residual solvent, none of which is visible chromatographically.
Discrepancies of one to two percentage points between a supplier’s figure and an independent determination are common and usually explicable: a faster gradient resolves fewer impurities, a higher integration threshold discards more small peaks, and a wider solvent-front exclusion window removes early-eluting fragments from the calculation altogether. None of that is misconduct. All of it is invisible unless the method is disclosed.
"The useful thing in that report is not the percentage, it is the system suitability data," said Rowena Ashcombe, community pharmacist and diabetes educator. "It is the only part that tells you whether the percentage means anything."
ERP’s dossier has been updated with this determination and with the date of the request for comment. The dossier’s "what we could not verify" box records the questions still outstanding.
Accreditation to the international competence standard covers the scope a laboratory has been assessed for, which is not necessarily the test you commissioned.
Reported from the analysis, not from a warning notice.
A compendial sterility test takes fourteen days of incubation and consumes the containers tested. That timescale is incompatible with the business model that made…
Reported from the analysis, not from a warning notice.
The supplier has not disputed the finding. It has not explained the gap either.
The supplier has not disputed the finding. It has not explained the gap either.