Set the threshold at a tenth of a per cent and the small impurities vanish

Baseline placement under a tailing peak is a judgement call with a direct effect on the area. It is one of the few places where two competent analysts genuinely disagree.

The threshold is the parameter with the largest effect. Below it a feature is treated as noise and does not enter the calculation; above it, it becomes an impurity subtracting from the reported purity. A threshold at a tenth of one per cent of the main peak area discards a population of small related species that a threshold at two-hundredths of a per cent would report individually. Both settings are defensible for their purposes. Only one of them is capable of producing a purity figure below 99 on a well-made peptide, which is a fact about the parameter rather than about the powder.

Gradient slope, expressed properly

A gradient should be quoted as a rate, not as a duration. Twenty-five to forty-five per cent acetonitrile over forty minutes is half a percentage point of organic per minute. The same range in twelve minutes is about one and two-thirds points per minute. That threefold difference in slope is the difference between resolving a deamidated relative from its parent and delivering both as one peak.

The underlying relationship is well established in peptide chromatography: resolution of closely related species improves as gradient slope decreases, up to the point where peak broadening from extended run times starts to give the gain back. Peak capacity — the number of peaks a method can theoretically resolve across its run — rises with shallower gradients and with more efficient columns, and it is the honest single-number summary of what a separation can do. It is never quoted in this trade.

The Journal’s standing request is simply that the gradient be printed. It is three numbers: starting composition, ending composition, time. Nobody regards it as commercially sensitive, every laboratory has it in the method file, and its presence converts a purity figure from an assertion into something comparable with the next certificate. Its absence is the reason two figures from two suppliers cannot be placed side by side, and that absence is a documentary decision rather than a technical constraint.1

The threshold, and what falls below it

Integration software applies a threshold — expressed as a slope sensitivity, an area cut-off, a height cut-off or some combination — below which a feature in the trace is treated as baseline noise and not integrated. The setting is necessary: without it, every fluctuation would be reported as an impurity and the result would be dominated by noise. The setting is also consequential, because a great many real, small, closely related species live in the region between the two conventional choices.

The arithmetic is easy to underestimate. A well-made peptide preparation may carry twenty or thirty related species each between two-hundredths and a tenth of one per cent — deletion sequences, deamidated and oxidised forms, epimers. Reported individually against a low threshold they might total a percentage point or more. Discarded against a high threshold they total zero. Two laboratories reporting 99.4 and 98.3 on the same lot may have measured the same chromatogram and disagreed only about which features are noise.

The Journal has asked all four independent services what threshold their standard peptide report uses. Two answered with a figure. One answered that it depends on the method and offered to supply the value per report, which is a better answer than a fixed number. One did not answer. We regard the threshold as second only to the gradient in importance and, like the gradient, it is a single value that whoever produced the document already knows.

There are legitimate ways to raise a purity figure that involve no change whatever to the material.

On competing over a single metric

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.2

One sample, four method conditions, four purity figures
ConditionGradient rate (%ACN/min)Run time (min)ThresholdPurity reported
A1.67120.10%99.3%
B1.67120.05%98.9%
C0.50400.10%98.4%
D0.50400.05%97.5%
One physical sample from one vial, one instrument, one analyst, one afternoon; 25–45% acetonitrile in both gradients, 214 nm, identical column and injection load. The 1.8-point spread is attributable entirely to gradient slope and integration threshold. The conditions were specified by the Journal and do not represent the standard practice of the laboratory concerned.

Documentation practice, named

Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.

Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.

The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.

Dissolution, pH and what happens before the injection

A peptide in a vial exists in whatever state the manufacturer left it. Reconstituted in water, a hydrophobic sequence may not dissolve completely and the chromatogram will show particles or aggregates. Reconstituted in an acidic buffer, the same sequence dissolves and the chromatogram shows monomer. The purity figure—and the mass balance—changes accordingly. Sample preparation is not one of the twelve values that belong on a method disclosure, and yet it is one of the most consequential, because it determines what population the peptide is actually in when the injection happens.3

The trade addresses this by assuming that samples are dissolved in the mobile phase or its aqueous component, and the assumption is sometimes true and sometimes false. A certificate that states how the sample was dissolved, at what concentration, in what solvent and after what incubation time, is one that can be repeated. A certificate that does not is one that will produce different results if the receiving laboratory uses different dissolution practice, which is particularly consequential for poorly soluble sequences.

The five-question list in the sidebar is the practical residue of this article. Gradient, wavelength, threshold, standard, second method. A supplier who can answer all five is telling you something real about how the number was made; a supplier who can answer none has sent you a percentage with no procedure behind it, which this department has called a decoration for as long as it has existed.

References

  1. “Gradient slope, peak capacity and the resolution of closely related peptide impurities.” Journal of Chromatography A. 2015;1421:39–52.
  2. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  3. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.