Size exclusion sees the thing reversed-phase dissolved

Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.

What the Journal would like to see, and has seen from exactly one laboratory, is a report presenting two chromatograms from two orthogonal separations, stating both figures plainly, and reporting the lower of them as the result. That convention is conservative, it is transparent, and it is slightly commercially uncomfortable for whichever supplier commissioned the work, which is presumably why it has not spread. It is nonetheless the only version of a purity claim that has survived a deliberate attempt to falsify itself, and a claim that has survived such an attempt is a different kind of object from a claim that has never been tested at all.

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.1

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.2

Retention-time agreement is consistency. Molecular mass is composition. Only fragmentation approaches sequence.

On three claims that share one phrase

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

Method disclosure across twenty companies’ standard certificates
Disclosed itemOn standard certificateOn requestNot available
Purity figure2000
Method named as HPLC2000
Detection wavelength569
Gradient programme or rate1514
Integration threshold2315
Solvent-front exclusion window0218
Three largest impurities listed1118
Chromatogram attached479
Compiled from standard release documentation and from a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. Where a supplier attaches an independent laboratory report rather than transcribing a figure, the disclosure is credited to the certificate.

Documentation practice, named

Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.

Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.

The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

112815120-11Steep, 1.67 %/minShallow, 0.50 %/min0.860.90.940.9711.031.071.121.18normalised retention (parent = 1.00)relative response
Figure. Modelled detector traces for the same mixture under a steep and a shallow gradient, plotted against normalised retention. Two impurities resolved by the shallow gradient are carried into the parent peak by the steep one.

The full spectrum, recorded and discarded

A diode-array detector records ultraviolet absorbance across a wavelength range—commonly 200 to 400 nanometres—at every point in the chromatogram. From that data, two things of immediate value can be extracted. First, the spectrum can be examined at each peak to confirm spectral homogeneity—a pure peak has a spectrum that does not drift across its width, while a co-eluted pair shows a spectrum that changes. Second, the full spectrum can be plotted to reveal absorbers that were not visible at the single monitoring wavelength.3 A 254-nanometre impurity in a sample monitored at 280 nanometres is invisible in the 214-nanometre chromatogram and is therefore invisible in the purity report, until a diode-array scan reveals it.

Almost no research-peptide certificate prints this information. The data exists—it is generated automatically—and its absence from the report is a formatting decision, not a technical limitation. The usefulness of seeing the full spectrum would be apparent on the first certificate where it resolves an otherwise inexplicable discrepancy between the purity and the mass-balance calculation. It is one of the few changes to the standard format that costs nothing and adds information the reader genuinely needs.

Dissolution, pH and what happens before the injection

A peptide in a vial exists in whatever state the manufacturer left it. Reconstituted in water, a hydrophobic sequence may not dissolve completely and the chromatogram will show particles or aggregates. Reconstituted in an acidic buffer, the same sequence dissolves and the chromatogram shows monomer. The purity figure—and the mass balance—changes accordingly. Sample preparation is not one of the twelve values that belong on a method disclosure, and yet it is one of the most consequential, because it determines what population the peptide is actually in when the injection happens.4

The trade addresses this by assuming that samples are dissolved in the mobile phase or its aqueous component, and the assumption is sometimes true and sometimes false. A certificate that states how the sample was dissolved, at what concentration, in what solvent and after what incubation time, is one that can be repeated. A certificate that does not is one that will produce different results if the receiving laboratory uses different dissolution practice, which is particularly consequential for poorly soluble sequences.

Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.

References

  1. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  2. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.
  3. European Pharmacopoeia. Chapter 2.2.29 — Liquid Chromatography. Council of Europe, Strasbourg.
  4. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.

Letters to the Editor

5 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Your article argues for two orthogonal methods and reports the lower figure, but the people running a single twelve-minute method have a cost story you do not address. A full orthogonal pair doubles the turnaround and at least doubles the cost, which is why the market does not do it. The criticism of method disclosure is fair. The criticism that a single method is wrong is unfair to the constraints people operate under.

R. Whitlam, Adelaide, SA

The Journal replies

We are careful to say that a second method costs instrument time on a sample already in the autosampler, which is substantially less than twice the turnaround, but you are right that we underweight the commercial reality that a buyer setting a budget for testing is trading thoroughness for speed and price. Where we would push back is that those constraints are not technical or regulatory ones. They are market ones, and markets can change if enough buyers demand it.

Sub-two-micron columns changed everything about what is practical for peptide separations, but I would push back on the statement that particle size gains are costless. The pressure limit of most commercial instruments is three hundred bar, and trying to force 1.7-micron particles at eight millilitres per minute on a 4.6-millimetre column will send you there quickly. Peak capacity is not free.

C. Rautenbach, Pretoria

The Journal replies

Correct on the pressure cost, and that belongs in the method section. The trade-off is real, which is why many laboratories use core-shell superficially porous particles as a compromise: they are substantially cheaper than sub-two-micron packings, deliver most of the efficiency gain at a lower pressure, and the efficiency-cost-pressure triangle is the actual business decision people make. We should have named it.

One practical note from the other side of the counter. When a customer asks us for the gradient we send it. When a customer asks a reseller, the reseller does not have it, because they were sent a PDF with a number on it. The gap you are describing is often two links down the chain rather than at the laboratory.

D. Mazzarella, Catania

The Journal replies

That is an important structural point and it changes where the fix has to happen. If the laboratory report travels intact instead of being transcribed into a house certificate, the method travels with it. Four of the twenty companies we track already attach the original report, and on this argument they are doing the single most useful thing available.

Your table of what each method can see puts "only if resolved" against isoaspartate for RP-HPLC. That understates the difficulty. Resolving isoAsp from Asp routinely requires a method developed for the purpose, and on a generic gradient the two are frequently indistinguishable even at forty minutes.

A. Mbeki, Lusaka

The Journal replies

Accepted, and the entry now reads that it requires a method developed for the purpose. Our original wording implied that a sufficiently shallow generic gradient would generally do it, which overstates what shallowness alone achieves.

You list five things a purity figure cannot tell you and then say the list is not an indictment of the technique. It reads like one. If a measurement is silent on content, aggregation, sequence, isomers and microbiology, why is it the measurement this market uses at all?

C. Adeoti, Ibadan

The Journal replies

Because it is cheap, fast, comparable-looking and genuinely informative about the thing it measures. A tyre pressure gauge is silent on tread depth, brake pads and the driver, and it is still the right instrument for its question. The failure is in a market that owns one gauge and calls the reading roadworthiness.