The empty half of every certificate of analysis

Purity, content, identity, sterility and endotoxin are five separate determinations. A single sheet of paper carrying one of them is not evidence about the other four.

A certificate of analysis for a research peptide is, in the overwhelming majority of cases, a document about chemistry. It reports a chromatographic purity figure, sometimes a mass, occasionally a water content, and it is generated from a few milligrams of powder dissolved in a vial of mobile phase. Nothing in that procedure has any bearing on whether the sealed container it came from holds viable microorganisms, whether the powder carries pyrogenic material from an upstream water system, or whether the stopper has maintained a seal since the day it was crimped. The document is accurate and it is answering a different question.

Two questions that share a piece of paper

Consider what happens physically when a certificate is produced. A few milligrams of lyophilised powder are weighed, dissolved in an aqueous mobile phase with an organic modifier, injected onto a reversed-phase column and separated over a programmed gradient while an ultraviolet detector records absorbance. The output is a trace. Software integrates the areas beneath its features, and the main peak area as a proportion of the total becomes the purity figure.

Every step of that procedure is blind to microbial contamination. A bacterial cell contributes no ultraviolet-absorbing peak at any retention time a peptide method would record. Endotoxin, a lipopolysaccharide, is not usefully detected at the wavelengths used for peptide bond absorbance and would in any case be present at a mass fraction several orders of magnitude below any integration threshold in commercial use. A vial holding a hundred colony-forming units and a vial holding none produce chromatograms that no analyst could distinguish.

This is not a defect of the method. Reversed-phase chromatography is an excellent way of determining what proportion of the chromatographically visible material is the intended species, and that is what it is being asked. The defect is in the reading. A document answering one question is being filed as evidence about five.

Depyrogenation, and the step that gets skipped

Because endotoxin survives sterilisation, glass components require a separate treatment. The standard is dry heat: a tunnel or oven cycle at two hundred and fifty degrees or above, validated to achieve at least a three-log reduction in a deliberately applied endotoxin challenge. Vials emerging from a qualified depyrogenation tunnel are both sterile and pyrogen-free, and they remain so only if the subsequent handling maintains it.

Elastomeric closures cannot take that treatment and are instead washed, siliconised and steam-sterilised by a validated cycle, with endotoxin control achieved by the washing step and specified as a limit per stopper. Water for injection is controlled at source, with a compendial endotoxin specification, and a water system is the commonest origin of a pyrogen problem in an otherwise competent facility.

Depyrogenation is also, in the Journal’s experience of asking, the step most frequently absent from descriptions of small fill operations. Autoclaved vials are sterile. Autoclaved vials are not depyrogenated, and an operation that describes its glass preparation solely in terms of autoclaving has told you something specific about what it has not done. We put this point to eleven correspondents; two said their glass is depyrogenated by dry heat with a validated cycle, and one asked us what depyrogenation was, which we took as a straight answer and a useful one.

A dye ingress test can pass a container with a defect large enough to admit an organism, because the dye happened not to travel.

On probabilistic leak testing

What we submitted, and what came back

Between the second and fourth quarters the Journal purchased fourteen vials of lyophilised research peptide from nine suppliers, at catalogue prices, through ordinary channels and without identifying ourselves. Each was photographed sealed, logged, and stored at two to eight degrees on arrival. Twelve were submitted to a contract laboratory accredited to the general competence standard for testing laboratories for bacterial endotoxin determination by kinetic chromogenic assay, with method suitability established for each matrix. Two were submitted for a compendial sterility test by membrane filtration, which destroyed both.

We disclose the following limitations without being asked. Fourteen vials from nine suppliers is not a survey. Single determinations carry the uncertainty of single determinations. A negative sterility result on one vial says nothing about the batch it came from, for exactly the statistical reasons set out above. And a vial that has crossed a border in a padded envelope has a thermal and mechanical history we cannot reconstruct.

What the exercise establishes is narrower than a survey and, we think, worth publishing: that the tests exist, that they are commercially available to a private purchaser at a known price, that the numbers they return are interpretable against a calculable limit, and that nothing prevented any of the nine suppliers from commissioning them first.1

Container closure integrity methods, by philosophy
MethodClassNon-destructiveApproximate defect resolution
Dye ingress under vacuumProbabilisticNo10–20 µm, poorly characterised
Microbial immersion challengeProbabilisticNo5–10 µm, highly variable
Vacuum decayDeterministicYes2–5 µm
High-voltage leak detectionDeterministicYes1–5 µm, liquid fills
Laser headspace analysisDeterministicYesSub-micron, stability-programme capable
Helium mass spectrometryDeterministicYesSub-micron
Resolution figures are indicative and package-dependent; the meaningful specification for any given system is its maximum allowable leakage limit, established experimentally rather than assumed.

Documentation practice, named and criticised

The Journal tracks the release documentation of twenty companies. On sterility and endotoxin the picture is close to uniform. Every one publishes a purity figure. A minority publish an identity confirmation. Two publish peptide content. On the microbiological attributes, the standard document is silent, and the silence is not annotated: there is no line reading that sterility has not been determined, which would at least be informative.

Some of the practices we would like to see are already in use somewhere in the group. SSA reports peptide content on its certificates, having begun after correspondence with this publication. CPC and SWB describe their fill environment in general terms on request. QST and BCH answered our five questions in full. WXT and FGP declined on the grounds that research-use products are not represented as sterile injectables, which is a legally sound answer that concedes the point of the exercise. Several others did not respond, and we record non-response as non-response rather than as evasion.

What we are criticising is a documentary convention, not the conduct of any company named here. None of the twenty has been shown to us to have misrepresented anything. The convention is that a chemistry certificate stands in for a release package, and it is a convention this trade adopted collectively and could abandon the same way.

What this article is not asserting

Precision about the scope of a criticism is part of the job, so it is worth spending a paragraph on what is not being said. This piece does not allege that any company named in it has sold contaminated material. It does not allege that any of them has concealed a result, falsified a document or misrepresented a test. It does not claim that the products discussed are dangerous, and it does not claim that they are safe, because neither claim is supportable from the evidence we have.

What the piece asserts is documentary. A certificate describing chemistry is being read as a release package covering microbiology. That mismatch is created by the format of the document rather than by anybody’s intent, and it is closed by adding lines rather than by changing behaviour.

There is also a legal point the Journal has no wish to elide. Research-use-only material is not approved for human use in any jurisdiction, is not required to meet parenteral standards, and is not represented by its sellers as meeting them. Everything in this article about endotoxin limits and sterility assurance describes the framework that would apply to a parenteral medicine. Applying that framework to a research chemical is a comparison, not a compliance requirement, and readers should hold both halves of that sentence at once.

A note on method and sourcing

The compendial material in this article is drawn from the current general chapters of the United States Pharmacopeia and the European Pharmacopoeia, read in the original rather than in summary, and from the international standards on aseptic processing and on laboratory competence. Where a chapter has changed status recently — as the recombinant reagent chapters have — we say so, because a reader consulting an older edition will find a different framing.

Where the Journal reports a number it obtained itself, it states the laboratory’s accreditation status, the method family, whether method suitability was established, and the number of determinations. Where we report what a company told us, we distinguish an answer from a refusal and a refusal from a non-response, because those three things are routinely collapsed in coverage of this trade and they are not the same.

Corrections to this department are handled by the standards desk, which reads every letter and records the outcome in the log. Readers who believe a paragraph here overstates its evidence are asked to write to standards@compoundjournal.com; readers with documents to send, including certificates they would like read, should write to letters@compoundjournal.com. We do not publish correspondents’ names without permission and we do not identify the source of a certificate.

We will keep asking the five questions, printing the answers, and recording the refusals as refusals. Two companies changed their certificates after the first round of this correspondence, which is a small result for a year of letters and rather better than none. Documents to letters@compoundjournal.com; disputes about anything above to standards@compoundjournal.com.