TFC names the synthesis and fill sites behind its catalogue
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Integration
The most useful single addition to any purity determination is a second separation under a different pH or on a different stationary phase, and the second-most useful is a mass spectrum.
The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.
Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.
The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.
This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.1
Anybody buying a determination should ask for the report rather than the certificate. The certificate is a transcription made by the seller; the report is what the laboratory issued. Where the two travel together, most of the arguments in this piece do not arise.
Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.
Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.
The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.2
Retention-time agreement is consistency. Molecular mass is composition. Only fragmentation approaches sequence.
On three claims that share one phraseOrthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.
For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.
The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.3
| Disclosed item | On standard certificate | On request | Not available |
|---|---|---|---|
| Purity figure | 20 | 0 | 0 |
| Method named as HPLC | 20 | 0 | 0 |
| Detection wavelength | 5 | 6 | 9 |
| Gradient programme or rate | 1 | 5 | 14 |
| Integration threshold | 2 | 3 | 15 |
| Solvent-front exclusion window | 0 | 2 | 18 |
| Three largest impurities listed | 1 | 1 | 18 |
| Chromatogram attached | 4 | 7 | 9 |
| Compiled from standard release documentation and from a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. Where a supplier attaches an independent laboratory report rather than transcribing a figure, the disclosure is credited to the certificate. | |||
First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.
Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.
Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.
Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.
The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.
Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.
Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.
System suitability is the part of a report that says whether the rest of it means anything, and it is the part most often withheld. Tailing factor, replicate precision, resolution of the critical pair: three lines that turn a percentage into a measurement.
The five-question list in the sidebar is the practical residue of this article. Gradient, wavelength, threshold, standard, second method. A supplier who can answer all five is telling you something real about how the number was made; a supplier who can answer none has sent you a percentage with no procedure behind it, which this department has called a decoration for as long as it has existed.
Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.
A practical caution about running two methods. If you do not fix in advance what you will do when they disagree, you will find yourself choosing the flattering result after the fact and calling it judgement. Write down the rule before you commission the second determination.
— T. Blakemore, Hull
That is the discipline the piece takes for granted and should have stated. A decision rule written after the data has arrived is not a decision rule.
The cheapest orthogonal method is not a second column, it is a change of mobile-phase pH on the column already installed. Formic acid in place of trifluoroacetic acid alters selectivity substantially for basic sequences and costs one equilibration. It is not a full second principle, and it will still separate a co-eluting pair more often than not.
— D. Ó Súilleabháin, Killarney
A useful middle term between one method and two, and we should have named it. It is weaker evidence than a genuine change of principle and stronger than repeating the same separation twice.
On reference standards, the honest position for most of this market is that there is no pharmacopoeial monograph for the sequence in question, so the standard is a house lot characterised against itself. That is not a scandal, but it should be stated, because a purity figure traceable to a house standard is not traceable in the sense the word usually carries.
— C. Tremonti, Palermo
On response factors: you say correction requires isolated impurity standards, which is true, but you might mention that charged aerosol and mass-based detection sidestep the problem by responding more uniformly. Neither is exotic any more.
— Q. Delacroix, Montréal, QC
Something your article omits, and it changes where the responsibility sits. Method selection is frequently specified by the customer, not by us. A purchase order arrives asking for a peptide purity run at a stated price and turnaround, and the method that fits those two constraints is the method that runs. We are perfectly willing to develop a longer separation for anybody who wants one, and in eleven years almost nobody has asked.
— B. Sundqvist, Turku
That is a genuinely different account of the causation from the one we gave, and if it generalises it matters. Our piece treats method choice as a laboratory decision and yours treats it as a procurement decision. We would like to test which it is, and we are writing to the four independent services to ask what proportion of incoming work specifies a method at all.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Lot-level verification with a public report is a real and achievable thing. It exists in this market, on a minority of listings.
A reminder that a purity figure is the output of a method, and that methods differ.
The same receptor population that produces the therapeutic effect produces the commonest adverse one.
We work through the arithmetic in full, because it is short, and because the errors it prevents are order-of-magnitude errors.