Janoshik result on a Nantong exenatide lot lands 3.1 points below the supplier’s figure
The Journal submitted the sample and paid for the analysis. The vendor was told in advance.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Analytics
Regulated pharmaceutical practice has explicit reporting, identification and qualification thresholds. This trade has none, which is why a purity figure carries no information about what was disregarded.
The threshold is the parameter with the largest effect. Below it a feature is treated as noise and does not enter the calculation; above it, it becomes an impurity subtracting from the reported purity. A threshold at a tenth of one per cent of the main peak area discards a population of small related species that a threshold at two-hundredths of a per cent would report individually. Both settings are defensible for their purposes. Only one of them is capable of producing a purity figure below 99 on a well-made peptide, which is a fact about the parameter rather than about the powder.
A gradient should be quoted as a rate, not as a duration. Twenty-five to forty-five per cent acetonitrile over forty minutes is half a percentage point of organic per minute. The same range in twelve minutes is about one and two-thirds points per minute. That threefold difference in slope is the difference between resolving a deamidated relative from its parent and delivering both as one peak.
The underlying relationship is well established in peptide chromatography: resolution of closely related species improves as gradient slope decreases, up to the point where peak broadening from extended run times starts to give the gain back. Peak capacity — the number of peaks a method can theoretically resolve across its run — rises with shallower gradients and with more efficient columns, and it is the honest single-number summary of what a separation can do. It is never quoted in this trade.
The Journal’s standing request is simply that the gradient be printed. It is three numbers: starting composition, ending composition, time. Nobody regards it as commercially sensitive, every laboratory has it in the method file, and its presence converts a purity figure from an assertion into something comparable with the next certificate. Its absence is the reason two figures from two suppliers cannot be placed side by side, and that absence is a documentary decision rather than a technical constraint.1
Under a tailing peak, the area depends on where the baseline is drawn, and the software offers several conventions — a straight line between valley points, an exponential skim, a tangential skim for a shoulder. For a symmetrical peak on a flat baseline the differences are trivial. For a small impurity riding on the tail of a large parent they are not, and they are the commonest genuine disagreement between two competent analysts looking at the same trace.
The denominator is decided by the solvent-front convention. Everything eluting in the first moments of a gradient run — unretained salts, injection solvent, dissolved gases, mobile-phase impurities — is conventionally excluded, and rightly so. Methods differ on where the exclusion window ends, and the choice determines whether an early-eluting hydrophilic fragment is an impurity or a non-event. The difference between a window closing at one and a quarter minutes and one closing at two and a half is not analytical laxity; it is a judgement about the separation, and it moves the number.
Manual reintegration deserves a word, because it has an undeserved reputation. Reintegrating by hand is entirely legitimate — software misassigns peak boundaries regularly — and in regulated practice it is permitted, documented and audited, with the original and revised integrations retained. What is not acceptable is undeclared reintegration, and in this market there is no mechanism by which it would ever be declared.2
Resolution is the joint product of efficiency and selectivity. Improvement in one does not compensate for inadequacy in the other.
On the method trade-offThe Journal buys material and has it tested, and the design of those exercises deserves the same disclosure we ask of others. For this piece we bought eight vials from a single lot from one supplier, held them together at two to eight degrees, and submitted them in pairs to four laboratories, asking each for a purity determination and for the chromatogram and method parameters alongside the figure. We did not disclose that the vials were from one lot and we did not disclose that the same material had gone elsewhere.
Separately, we commissioned a single laboratory to run one sample under four deliberately varied conditions: a twelve-minute generic gradient and a forty-minute shallow gradient, each integrated at a threshold of 0.1 per cent and 0.05 per cent. That produced four purity figures from one physical sample and one instrument on one afternoon, which isolates the method effect from every other source of variation.
The limitations are ours to state. One lot from one supplier is not a survey of the market. Single injections carry the variability of single injections. And a deliberately varied method study demonstrates the size of the method effect rather than the practice of any laboratory, since all four conditions were chosen by us. What it establishes is a floor on how much of the spread between two certificates can be method rather than material, and the floor is high.3
| Condition | Gradient rate (%ACN/min) | Run time (min) | Threshold | Purity reported |
|---|---|---|---|---|
| A | 1.67 | 12 | 0.10% | 99.3% |
| B | 1.67 | 12 | 0.05% | 98.9% |
| C | 0.50 | 40 | 0.10% | 98.4% |
| D | 0.50 | 40 | 0.05% | 97.5% |
| One physical sample from one vial, one instrument, one analyst, one afternoon; 25–45% acetonitrile in both gradients, 214 nm, identical column and injection load. The 1.8-point spread is attributable entirely to gradient slope and integration threshold. The conditions were specified by the Journal and do not represent the standard practice of the laboratory concerned. | ||||
Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.
Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.
The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.
A peptide in a vial exists in whatever state the manufacturer left it. Reconstituted in water, a hydrophobic sequence may not dissolve completely and the chromatogram will show particles or aggregates. Reconstituted in an acidic buffer, the same sequence dissolves and the chromatogram shows monomer. The purity figure—and the mass balance—changes accordingly. Sample preparation is not one of the twelve values that belong on a method disclosure, and yet it is one of the most consequential, because it determines what population the peptide is actually in when the injection happens.4
The trade addresses this by assuming that samples are dissolved in the mobile phase or its aqueous component, and the assumption is sometimes true and sometimes false. A certificate that states how the sample was dissolved, at what concentration, in what solvent and after what incubation time, is one that can be repeated. A certificate that does not is one that will produce different results if the receiving laboratory uses different dissolution practice, which is particularly consequential for poorly soluble sequences.
The technique is not on trial here and never was. Reversed-phase chromatography can resolve species differing by a single methyl group, and the laboratories running it for this market are, on the evidence we have gathered, largely competent and entirely willing to describe what they did when somebody asks. What is on trial is a document format that omits the four values needed to compare one figure with another.
The Journal submitted the sample and paid for the analysis. The vendor was told in advance.
This piece takes the measurement apart into the decisions it is made of, because each decision moves the answer.
Multiple-dose vials are designed for a defined number of punctures. Nobody counts, and the elastomer does not care whether anybody counts.
Every step between the laboratory report and the product page removes information, and the badge is the last step.
The supplier has not disputed the finding. It has not explained the gap either.
Two years ago we ran an anonymised version of this comparison and promised a named one. This is it, with every method printed in full.