Why PeptideMeter may run electrospray where another laboratory runs MALDI

A brief and unromantic tour of the interface between a liquid sample and a vacuum.

A mass spectrometer cannot weigh a peptide. It can only measure the trajectory of a charged particle through electric and magnetic fields, from which a mass-to-charge ratio is calculated. Everything reported on a certificate as a molecular weight is therefore a two-step inference: an ion was observed at a particular mass-to-charge ratio, and a charge was assigned to it. Both steps can go wrong, and both are omitted from almost every identity statement circulating in this trade. The first question to ask of any mass figure is not whether it matches. It is how the molecule was charged.

The instrument does not weigh anything

It is worth being exact about what a mass spectrometer does, because the imprecision propagates. The instrument generates ions from a sample, separates them according to the ratio of their mass to their charge, and counts them at a detector. The horizontal axis of every spectrum is mass-to-charge, conventionally written m/z and expressed in thomsons or in dimensionless units depending on the vendor’s software. Nothing is weighed. Nothing is measured against a reference mass in the sense that a balance measures against a calibration weight.

What follows from this is that every molecular weight on every certificate of analysis in this market is a calculated quantity, derived from a measured m/z by assigning a charge and subtracting the mass contribution of whatever adducted to the molecule to give it that charge — usually protons, sometimes sodium, occasionally potassium or ammonium. The assignment is normally straightforward and normally correct. It is nonetheless an assignment, and when it goes wrong it goes wrong by an integer factor, which is the kind of error that produces confident nonsense rather than a plausible discrepancy.

The practical consequence for a reader is a habit: when a mass figure appears, ask what was observed and what was inferred. A report that gives both — the m/z, the charge, and the derived neutral mass — has answered the question before it was asked.

Electrospray, and why it produces a family of ions

Electrospray ionisation works by pumping a solution of the analyte through a fine capillary held at a potential of a few kilovolts relative to the instrument’s entrance. The liquid emerging from the tip forms a cone and then a jet of charged droplets. As solvent evaporates the droplets shrink, the charge density on their surfaces rises, and at the point where electrostatic repulsion exceeds surface tension they fission into smaller droplets. Repeat this enough times and what is left is a bare, charged analyte ion in the gas phase.

Because the charge is acquired in solution and retained through desolvation, a peptide with several basic residues will carry several protons, and the population of ions reaching the analyser is distributed across charge states. This is the defining characteristic of electrospray spectra and the reason they look bewildering to a first-time reader: a single pure compound produces four or five prominent peaks, none of them at the molecular weight.

The distribution is not noise. It carries information about the number of accessible basic sites and about the conformational state of the molecule in solution, and it shifts predictably with mobile-phase composition and pH. It also has a practical advantage that matters for identity work: dividing the mass by three or four brings a large peptide into the range where instruments achieve their best resolving power and accuracy.1

A tolerance that no plausible measurement could fail is not an acceptance criterion. It is a formality.

On the ±1 dalton convention

MALDI, the matrix, and the singly charged ion

Matrix-assisted laser desorption ionisation takes a different route. The peptide is mixed with a large molar excess of a small organic compound that absorbs strongly at the laser wavelength — α-cyano-4-hydroxycinnamic acid and sinapinic acid are the usual choices for peptides and proteins respectively — and the mixture is dried on a metal target. A pulsed ultraviolet laser strikes the crystal, the matrix absorbs the energy, and a plume of matrix and analyte is ejected into the vacuum with the analyte largely intact and mostly singly protonated.

Two consequences follow. First, MALDI spectra are simple: one predominant ion per compound, at the molecular weight plus one proton, which makes them easy to read and easy to print on a certificate. Second, MALDI is markedly more tolerant of salts, buffers and heterogeneous samples than electrospray, which is why it survives in routine synthesis monitoring where electrospray would require a chromatographic clean-up first.

The trade-offs are equally real. The matrix produces intense chemical background below roughly a thousand daltons, which obscures small fragments. Ion yield varies between compounds and between spots on the same target, making MALDI a poor quantitative technique. And the achievable mass accuracy on a linear instrument at peptide molecular weights is measured in hundreds of parts per million unless a reflectron and delayed extraction are in use.2

Monoisotopic and average mass for peptides commonly encountered in this market
CompoundMonoisotopic (Da)Average (Da)Difference (Da)Difference (ppm)
BPC-1571418.681419.530.85600
Liraglutide3748.053751.203.15840
Semaglutide4111.124113.582.46598
Tirzepatide4810.474813.452.98619
Retatrutide4728.424731.302.88609
Tesamorelin5131.635135.904.27831
Values calculated from published molecular formulae for the free-base forms and rounded to two decimal places; salt forms and acylation variants shift these figures. The final column shows why a certificate that does not state its convention cannot be checked: the convention difference alone exceeds any plausible acceptance tolerance.

Which source a laboratory chooses, and why it should say so

The choice between the two techniques is not a matter of quality but of question. A synthesis chemist watching a coupling proceed wants a fast, salt-tolerant check that the chain has grown by the expected residue, and MALDI on a bench instrument answers that in minutes. An analytical laboratory asked whether a submitted vial contains the labelled compound and nothing closely related to it needs the resolving power and the accuracy that electrospray into a high-field analyser provides, coupled to a chromatographic separation so that species which co-elute can at least be assigned to retention times.

Both appear in this market, and reports rarely distinguish them. That matters because the two techniques have different blind spots. MALDI can induce loss of labile modifications during desorption, so a phosphorylated or otherwise fragile species may be under-represented. Electrospray suppresses ionisation of some analytes in the presence of others, so a minor component of a mixture may be absent from a spectrum in which it is genuinely present.

A certificate stating the source therefore tells a reader which class of error to consider. The Journal has stopped asking suppliers for more testing and started asking them for this line instead, on the grounds that it costs nothing and changes what the existing test can be said to support.

Oxidation: sixteen daltons, and the one that gets caught

Methionine oxidises to the sulphoxide with a mass increase of 15.995 daltons, and on to the sulphone at a further 15.995. Tryptophan and histidine oxidise by related routes. The chemistry is driven by dissolved oxygen, by trace peroxides in excipients and in some grades of polysorbate, by light, and by transition-metal contamination, and it proceeds in lyophilised material as well as in solution, though more slowly.

Analytically this is the easy case, and it deserves to be described as such in an article otherwise concerned with what cannot be seen. A sixteen-dalton shift is resolvable on essentially any instrument, and the oxidised species is usually chromatographically distinct enough to appear as a separate peak under a reasonably shallow gradient. A spectrum showing a plus-sixteen satellite at a few per cent of the parent intensity is unambiguous evidence of oxidation, and its absence is meaningful evidence of the opposite.

Which is why the Journal’s standing request to laboratories in this market is for the spectrum rather than the verdict. A reproduced spectrum, even at the modest resolution of a routine instrument, allows a reader to look for the plus-sixteen satellite themselves. A conformance statement does not, and cannot be made to. This is the cheapest available improvement to identity reporting in the trade and it consists of printing a picture the laboratory has already produced.

1097795492190-1130.98 Da shift0.036 Da shift1000200030004000500060008000peptide molecular weight (Da)ppm accuracy required
Figure. Mass accuracy required to resolve two shifts, as a function of peptide molecular weight. The upper series is deamidation at 0.98 Da; the lower is the glutamine-for-lysine exchange at 0.036 Da. The horizontal scale is where routine instruments sit: a few parts per million for an orbital trap, several hundred for a linear time-of-flight.

Six lines that would make an identity claim checkable

The Journal has settled on a short list, arrived at by writing to laboratories and asking what they could supply without additional work. Six lines. The ionisation source and mode. The analyser, named by class at minimum and by model preferably. The theoretical mass, with the convention stated as monoisotopic or average. The observed mass, with the charge state from which it was derived. The deviation, expressed in parts per million. And the acceptance criterion that was applied.

Every one of those is in front of the analyst at the moment the report is generated. None is commercially sensitive. Together they convert a verdict into a measurement, because they allow a reader to determine what the test could have detected and what it could not. A document carrying those six lines can be assessed by somebody who has never seen the sample; a document reading “MS: conforms” cannot be assessed at all, by anybody, including the person who wrote it.

The compendial approach to identity testing is built on the same three elements — a technique, a reference, and a criterion — and asks for them to be stated because a test whose discriminating power is undocumented has not been validated in any meaningful sense.3 We are not asking this market to become a regulated one. We are asking it to print what it already knows.

The standing rule in this department

This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.

The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.

What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.

Readers who take one thing from this piece should take the arithmetic. Isotope spacing is one over the charge. Deamidation is one dalton, oxidation is sixteen, and stereochemical inversion is nothing at all. A tolerance of ±1 dalton on a four-thousand-dalton peptide is two hundred and forty parts per million and excludes almost nothing worth excluding. Those four facts are sufficient to read most of the identity claims in circulation, and they fit on the back of an envelope.

References

  1. “Charge-state distributions in electrospray ionisation of peptides and their dependence on solution conditions.” Journal of the American Society for Mass Spectrometry. 2015;26(8):1319–1332.
  2. “Matrix selection, sample preparation and mass accuracy in MALDI time-of-flight analysis of synthetic peptides.” Rapid Communications in Mass Spectrometry. 2014;28(19):2077–2088.
  3. International Council for Harmonisation. Q2(R2): Validation of Analytical Procedures. 2023. Sections on specificity and on the demonstration of discriminating power.

Letters to the Editor

4 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

I would add one omission to your six lines: the date and nature of the last calibration. A parts-per-million figure from an instrument last calibrated a fortnight ago is a different claim from one calibrated that morning with an internal standard.

H. Baptiste, Fort-de-France

The Journal replies

Agreed, and it may be the best suggestion we have received on this subject. It is now a seventh line in the version of the list we send to suppliers, with the note that internal calibration should be stated where it was used.

Your article says a matching mass does not confirm a sequence, which is correct, and then rather implies that vendors are trading on the ambiguity. I run analytical services and I would put it differently: we report what we measured, in the words our clients ask for. If the Journal wants the word confirmed retired, write to the buyers, not to us.

R. Hollenbeck, Spokane, WA

The Journal replies

That is a fair reallocation of the criticism and we accept it. The word is chosen by whoever commissions the report, and laboratories are answering the question they were paid to answer. Our complaint is with the practice, not with the analysts, and the article should have located it more precisely.

On your point about D-amino acids: chiral amino-acid analysis after hydrolysis is not exotic and several contract laboratories offer it. The obstacle is that hydrolysis itself racemises a few per cent of most residues, so the method has a blank problem, and interpreting a low-level D content is genuinely difficult rather than merely expensive.

T. Elorriaga, San Sebastián

The Journal replies

An important qualification and we are glad to have it. The article implied the barrier was commercial when a substantial part of it is methodological. Recorded, and the section has been rewritten accordingly.

You state that fourteen of twenty suppliers report an MS identity test. Does that count reports supplied to you on request, or only what appears on the certificate a customer receives?

N. Halvorsen, Trondheim

The Journal replies

The former, which the table note now says explicitly. The count for what appears on a customer-facing certificate is lower in at least four cases, and we should have separated the two columns rather than merging them.