Duplicate submissions under different names test within-laboratory repeatability, which is a different quantity from between-laboratory reproducibility.
Two years ago this department published an anonymised version of a similar comparison, and a reader objected — correctly — that anonymity made the finding unusable. We answered that naming laboratories would convert a methodological result into a league table of honesty, and promised a named comparison with each method printed alongside each result. This is that piece. Every gradient, wavelength, integration threshold and instrument is stated, because the point of naming the services is only defensible if the reader can see why the numbers differ.
Janoshik Analytical is a Czech laboratory and the service whose reports circulate most widely in this market. Its core offering is chromatographic purity determination and peptide content by elemental nitrogen determination, with mass-spectrometric identity confirmation available, on samples submitted by post. Reports are issued to the submitter and carry the method, the gradient and integration detail, which places them at the more informative end of what this market produces.
Janoshik advertises in this publication. The relationship is disclosed by name on our funding page, no advertiser sees editorial copy before publication, and this section was written and edited under the same rules as every other in this department. Readers who consider the arrangement disqualifying are entitled to say so and several have; the correspondence is printed.
The Journal’s substantive observations are two. First, the practice of reporting content alongside purity is the single most useful thing any service in this market does, because content is the number that determines how much peptide a nominal mass represents and almost nobody else reports it. Second, the reports describe submitted samples, they say so, and the trade nonetheless quotes them as batch verification. That second point is a criticism of citation practice rather than of the laboratory, and it applies to every service in this article.
Medutest operates as a combined testing and verification service and does not advertise in this publication. Its assay work covers purity and identity on submitted material, and its verification arm addresses vendor documentation: whether a certificate exists for the lot on sale, whether the batch identifiers reconcile, and whether the supplier can produce the underlying laboratory report on request.
The combination is unusual and worth describing carefully, because it addresses a gap the assay-only services structurally cannot. An assay tells you what is in a vial. A documentation audit tells you whether the vendor can connect that vial to a record, which is the question this department spends most of its time on. Running both under one roof means a single report can note that the material is good and the paperwork is not, a combination the Journal encounters frequently and which no purity figure alone can express.
The corresponding difficulty is that a service performing both functions has two relationships with the same vendor, one of them adversarial. Medutest is not unique in this; VendorInvestigate faces a version of it. The Journal has no evidence of the conflict producing a bad outcome in any case it has examined, and notes it as a structural feature rather than a finding.
Within-laboratory repeatability was good in every case. Between-laboratory reproducibility was not, and every point of the gap is method.
On the blind duplicate exerciseTwelve vials, one lot, purchased at retail without disclosure of purpose. Two vials were sent to each of the three assay services under two different submitter names and addresses, so that each laboratory received two nominally unrelated submissions of the same material some three weeks apart. Six further vials were retained. Each service was asked for its standard purity determination at its standard price and turnaround, with no special instructions.
The design tests two distinct quantities that the trade conflates. Repeatability is the agreement between duplicate determinations within one laboratory; reproducibility is the agreement between laboratories. Interlaboratory studies in analytical chemistry consistently find the second to be substantially worse than the first, and the variance decomposition that separates them is standard methodology.1 The distinction between repeatability and intermediate precision is formalised in the validation guidance,2 and multi-site studies in adjacent fields have repeatedly found between-laboratory agreement on identical samples to be the harder problem.3
All three services were informed after the fact, before publication, and each was given the opportunity to comment on its own method as printed and on the comparison as a whole. All three responded. Two supplied additional method detail that has been incorporated. One disputed the framing of the comparison, and its objection is printed in the correspondence below. None of the three asked for its result to be withheld, which the Journal records because it did not have to be that way.
| Analysis | Quoted turnaround | Observed turnaround | Price band (EUR, single sample) |
|---|---|---|---|
| Purity, generic gradient | 3–5 working days | 4–9 days | 55–90 |
| Purity, extended gradient | 5–10 working days | 7–16 days | 110–180 |
| Purity + orthogonal confirmation | 2–3 weeks | 15–31 days | 190–320 |
| Identity by intact mass | 3–7 working days | 5–12 days | 45–110 |
| Peptide content by nitrogen | 1–2 weeks | 9–22 days | 160–280 |
| Water by Karl Fischer | 1 week | 6–11 days | 70–130 |
| Peptide mapping / sequence | 3–5 weeks | 26–38 days | 480–950 |
| Prices are the amounts actually invoiced to this publication at list rates between the second quarter of 2024 and the first quarter of 2026, converted where necessary at the rate on the invoice date, and are not quotations any reader should expect. Volume submitters pay materially less. Turnaround is measured from posting to receipt of the report. | |||
Within-laboratory repeatability was good. The two determinations from each service agreed to within 0.3 percentage points in every case, and to within 0.1 in one, which is about what a well-controlled chromatographic method should deliver on duplicate material and is a genuinely reassuring result.
Between-laboratory reproducibility was another matter. The three services returned figures spanning 2.1 percentage points on material from one lot. Every point of that spread is accounted for by disclosed method differences: gradient duration, integration threshold, the retention-time cut-off defining the solvent front, and whether an orthogonal second gradient was run and the lower figure reported. Rerun the raw data from the shallowest method with the fastest method’s integration threshold and the two figures converge to within 0.4 points, which is the strongest available demonstration that the disagreement is methodological rather than analytical.
Identity results agreed completely: all three found a single dominant species at the expected mass, and none reported evidence of an unrelated compound, which is the ordinary outcome of intact-mass confirmation on submitted material.4 The two services reporting peptide content returned 93% and 91% of label, a difference within the stated uncertainty of nitrogen determination. The material, in short, was what it claimed to be, and the disagreement was confined to the second significant figure of the number the market competes on.5
Four limitations, stated because the alternative is letting readers over-read a small study. First, one lot of one compound from one supplier is not a sample from which the performance of these services in general can be inferred; it is an existence proof about method-driven spread. Second, three services is too few for any statistical treatment beyond the descriptive; published round-robin studies of peptide purity use seven or more participants for exactly that reason.6
Third, and most important, the exercise tested reproducibility, not accuracy. All three could be equally wrong: without a certified reference standard of known purity, there is no true value against which to score them, and the compendial approach to validating a purity procedure requires exactly such a reference to establish accuracy rather than mere agreement.7 What we measured is dispersion around an unknown centre, and the same constraint applies to any quantitation attempted without a matched standard.8
Fourth, blind submission tests a laboratory’s ordinary process, which is the point, but it also means we bought the cheapest standard product from each service rather than the most thorough. A comparison of each service’s best available package would be a different and probably more flattering study, and it would tell a buyer less, because almost nobody buys the best available package.
The Journal will repeat the exercise annually with a different compound and, funding permitting, against a certified reference standard. The design is published so that others can run it.
Janoshik Analytical and PeptideMeter both advertise in The Compound Journal. Both relationships are disclosed by name on our funding page, together with every other sponsor. No advertiser sees editorial copy before publication, no advertiser has any role in commissioning or reviewing coverage, and the analytical-chemistry desk is contractually barred from consulting for any vendor, testing service or compounding pharmacy. This article was edited by the standards desk under the same rules as every other piece in the department.
The Journal also pays these services. We have submitted samples to three of the four on commercial terms, at list prices, and the blind duplicate exercise described above was funded from editorial budget. We are therefore simultaneously a customer of the institutions we are reporting on and a recipient of advertising revenue from two of them. Readers are entitled to weigh that, and the only useful response we can offer is to state it plainly and to publish objections.
Our position on the substance is unchanged by any of it. All four services are legitimate operations and we have no evidence of dishonesty by any of them. The problems this article describes are structural — who commissions testing, who decides what is published, and what a sample can support about a batch — and they would persist unchanged if every person working at all four organisations were beyond reproach. Correspondence to standards@compoundjournal.com.
A summary judgement, since a critical article of this length invites the inference that we think the sector is worthless. We do not. Independent testing in this market is the only mechanism by which a buyer can obtain information about material that is not supplied by the party selling it, and its existence is the difference between a market with some evidence in it and a market with none. Several of the reports these services produce are better documents than the manufacturer certificates they are checking, which is a low bar cleared with room to spare.
The three criticisms we would press are narrow. A sample is not a batch, and the trade cites samples as batches. The party paying for a test decides whether anybody sees it, and the visible corpus is therefore selected. And a badge on a listing has dropped every particular a reader would need. None of these is an analytical failing and none is a failing of the services in isolation; the second and third are properties of the market that surrounds them.
What we would tell a reader is this. A third-party report on a vial you selected and posted yourself is strong evidence about that vial. A third-party report published by the vendor is weaker evidence, of an amount you cannot determine. A badge is not evidence. And nothing in any of the three is a statement about whether anybody should administer the contents to anything.
One thing this article has deliberately not done is rank the four services. We do not think the evidence supports a ranking, we do not think a ranking would be used carefully, and two of the four advertise here, which is a reason for additional caution rather than a reason to pretend the relationship does not exist. What we have tried to publish instead is the material a reader would need to rank them for their own purposes.
Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.
I run analytical services and I object to the framing of your blind comparison. You bought our cheapest tier, published the number it produced alongside a competitor’s most thorough package, and called the result a spread. It is not a spread. It is three different products, priced accordingly, and your own table says so two columns to the right of the headline figure.
— J. Halloway, Dundee
This is the objection we thought was most likely and we think it is partly right. The comparison is of standard products at standard prices, which is what buyers actually purchase, and we said so. But the presentation invites the reading you object to, and the figure caption now states the tier alongside each result rather than leaving it to the method columns.
I submitted a vial to one of these services last year, got a result three points below what the vendor advertised, and did not know what to do with it. Your article explains why: I had one measurement on one vial, no method comparison, and no way to know if my vial was representative. I still do not know what to do with it, but I understand the shape of not knowing.
— K. Oyibo, Benin City
That is a better summary of this article’s practical content than our own closing paragraphs. The one thing we would add is that your result is worth publishing wherever you can, because buyer-submitted results are the scarcest and most informative category in the entire corpus.
The price table is the most useful thing you have published this year and also the thing most likely to be quoted out of context by somebody selling a comparison service. You might consider a note.
— K. Erdmann, Leipzig
There is one, and we have strengthened it. The figures are what this publication was invoiced at list rates and are not quotations a reader should expect; volume submitters pay materially less.
You note that no service offers sterility testing and that it takes fourteen days. Worth saying more plainly: a purity certificate and a sterility assurance are not merely different tests, they are different disciplines with different facilities, and no amount of chromatography will ever bear on it.
— Z. Karadzic, Novi Sad
Correct and worth the emphasis. We have said it in the certificates piece and should say it here too: nothing any of these four services sells addresses sterility, endotoxin or container closure integrity, and no combination of their reports adds up to one.
Reported from the analysis, not from a warning notice.
The report states the gradient, the wavelength and the integration threshold, which is more than most.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Efficacy was never the question in this appraisal. Duration of treatment was.
Efficacy was never the question in this appraisal. Duration of treatment was.