The two ionisation techniques in general use produce different charge distributions, different adducts and different failure modes. Certificates almost never say which was used.
A mass spectrometer cannot weigh a peptide. It can only measure the trajectory of a charged particle through electric and magnetic fields, from which a mass-to-charge ratio is calculated. Everything reported on a certificate as a molecular weight is therefore a two-step inference: an ion was observed at a particular mass-to-charge ratio, and a charge was assigned to it. Both steps can go wrong, and both are omitted from almost every identity statement circulating in this trade. The first question to ask of any mass figure is not whether it matches. It is how the molecule was charged.
It is worth being exact about what a mass spectrometer does, because the imprecision propagates. The instrument generates ions from a sample, separates them according to the ratio of their mass to their charge, and counts them at a detector. The horizontal axis of every spectrum is mass-to-charge, conventionally written m/z and expressed in thomsons or in dimensionless units depending on the vendor’s software. Nothing is weighed. Nothing is measured against a reference mass in the sense that a balance measures against a calibration weight.
What follows from this is that every molecular weight on every certificate of analysis in this market is a calculated quantity, derived from a measured m/z by assigning a charge and subtracting the mass contribution of whatever adducted to the molecule to give it that charge — usually protons, sometimes sodium, occasionally potassium or ammonium. The assignment is normally straightforward and normally correct. It is nonetheless an assignment, and when it goes wrong it goes wrong by an integer factor, which is the kind of error that produces confident nonsense rather than a plausible discrepancy.
The practical consequence for a reader is a habit: when a mass figure appears, ask what was observed and what was inferred. A report that gives both — the m/z, the charge, and the derived neutral mass — has answered the question before it was asked.
Electrospray ionisation works by pumping a solution of the analyte through a fine capillary held at a potential of a few kilovolts relative to the instrument’s entrance. The liquid emerging from the tip forms a cone and then a jet of charged droplets. As solvent evaporates the droplets shrink, the charge density on their surfaces rises, and at the point where electrostatic repulsion exceeds surface tension they fission into smaller droplets. Repeat this enough times and what is left is a bare, charged analyte ion in the gas phase.
Because the charge is acquired in solution and retained through desolvation, a peptide with several basic residues will carry several protons, and the population of ions reaching the analyser is distributed across charge states. This is the defining characteristic of electrospray spectra and the reason they look bewildering to a first-time reader: a single pure compound produces four or five prominent peaks, none of them at the molecular weight.
The distribution is not noise. It carries information about the number of accessible basic sites and about the conformational state of the molecule in solution, and it shifts predictably with mobile-phase composition and pH. It also has a practical advantage that matters for identity work: dividing the mass by three or four brings a large peptide into the range where instruments achieve their best resolving power and accuracy.1
Print the spectrum. It is the cheapest available improvement to identity reporting and it consists of reproducing a picture you already have.
The Journal’s request to laboratories, restated annuallyMatrix-assisted laser desorption ionisation takes a different route. The peptide is mixed with a large molar excess of a small organic compound that absorbs strongly at the laser wavelength — α-cyano-4-hydroxycinnamic acid and sinapinic acid are the usual choices for peptides and proteins respectively — and the mixture is dried on a metal target. A pulsed ultraviolet laser strikes the crystal, the matrix absorbs the energy, and a plume of matrix and analyte is ejected into the vacuum with the analyte largely intact and mostly singly protonated.
Two consequences follow. First, MALDI spectra are simple: one predominant ion per compound, at the molecular weight plus one proton, which makes them easy to read and easy to print on a certificate. Second, MALDI is markedly more tolerant of salts, buffers and heterogeneous samples than electrospray, which is why it survives in routine synthesis monitoring where electrospray would require a chromatographic clean-up first.
The trade-offs are equally real. The matrix produces intense chemical background below roughly a thousand daltons, which obscures small fragments. Ion yield varies between compounds and between spots on the same target, making MALDI a poor quantitative technique. And the achievable mass accuracy on a linear instrument at peptide molecular weights is measured in hundreds of parts per million unless a reflectron and delayed extraction are in use.2
| Analyser | Typical resolving power | Typical mass accuracy | Can assign charge from isotope spacing? |
|---|---|---|---|
| Single quadrupole | ~1,000 (unit) | 100–500 ppm | No |
| Linear ion trap | 2,000–4,000 | 50–200 ppm | At low m/z only |
| Linear MALDI-TOF | 500–1,500 | 200–1,000 ppm | No |
| Reflectron MALDI-TOF | 10,000–20,000 | 5–50 ppm | Yes |
| Quadrupole time-of-flight | 30,000–60,000 | 1–5 ppm | Yes |
| Orbital trap | 60,000–500,000 | <1–3 ppm | Yes |
| FT-ICR | >1,000,000 | <1 ppm | Yes |
| Figures are representative of instruments in general service and are quoted by manufacturers at favourable m/z values; performance at peptide molecular weights is generally lower. Accuracy figures assume routine calibration, and the better end of each range generally requires an internal calibrant. | |||
The choice between the two techniques is not a matter of quality but of question. A synthesis chemist watching a coupling proceed wants a fast, salt-tolerant check that the chain has grown by the expected residue, and MALDI on a bench instrument answers that in minutes. An analytical laboratory asked whether a submitted vial contains the labelled compound and nothing closely related to it needs the resolving power and the accuracy that electrospray into a high-field analyser provides, coupled to a chromatographic separation so that species which co-elute can at least be assigned to retention times.
Both appear in this market, and reports rarely distinguish them. That matters because the two techniques have different blind spots. MALDI can induce loss of labile modifications during desorption, so a phosphorylated or otherwise fragile species may be under-represented. Electrospray suppresses ionisation of some analytes in the presence of others, so a minor component of a mixture may be absent from a spectrum in which it is genuinely present.
A certificate stating the source therefore tells a reader which class of error to consider. The Journal has stopped asking suppliers for more testing and started asking them for this line instead, on the grounds that it costs nothing and changes what the existing test can be said to support.
Tandem mass spectrometry selects an ion of a particular mass-to-charge ratio, breaks it, and measures the masses of the pieces. Collision-induced dissociation and its higher-energy variant fragment the peptide predominantly at the amide bonds, producing two complementary series: b ions retaining the N-terminal portion and y ions retaining the C-terminal portion. The mass difference between consecutive members of either series is the residue mass of one amino acid, so reading the series in order reads the sequence.
Electron-transfer dissociation fragments differently, producing c and z ions, and preserves labile modifications that collisional methods tend to strip. Between them the two approaches cover most of what a peptide chemist needs. The nomenclature for these fragment series was fixed decades ago and is stable enough that a spectrum annotated in it can be read by anybody in the field.3
Two limitations should be stated because they are routinely elided. Fragmentation is not uniform along a chain: proline residues and basic residues bias cleavage, and stretches of a sequence can go unrepresented, which is why coverage is reported as a percentage rather than asserted as complete. And leucine and isoleucine remain indistinguishable under collisional fragmentation because their residue masses are identical; separating them requires side-chain fragmentation under specialised conditions, which almost nobody performs outside a research context.
This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.
The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.
What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.
Readers who take one thing from this piece should take the arithmetic. Isotope spacing is one over the charge. Deamidation is one dalton, oxidation is sixteen, and stereochemical inversion is nothing at all. A tolerance of ±1 dalton on a four-thousand-dalton peptide is two hundred and forty parts per million and excludes almost nothing worth excluding. Those four facts are sufficient to read most of the identity claims in circulation, and they fit on the back of an envelope.
Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.
You write that leucine and isoleucine cannot be distinguished by tandem mass spectrometry. That is too absolute. Side-chain fragmentation under high-energy conditions produces diagnostic w and d ions, and the discrimination has been demonstrated repeatedly.
— K. Sivertsen, Bergen
Correct, and the text has been amended. The discrimination is achievable under specialised conditions and is not available in any routine service this market uses, which is what we should have written rather than the stronger claim.
I have spent a week trying to reconcile a certificate’s stated mass of 4113.6 with a figure of 4111.1 I calculated from the sequence, and had convinced myself something was wrong with the vial. It was the isotope convention. Thank you, and also: how is this not stated on every certificate in existence?
— J. Vasilenko, Chisinau
We wish we knew. It is the single most common source of spurious discrepancies reaching this desk, it costs nothing to state, and we have now asked all twenty companies in the dossier programme to add it. Three have.
You give the glutamine-to-lysine difference as 0.036 daltons and say it needs an orbital trap. In practice you also need the two species to be chromatographically separated or present in a sensible ratio, because at one per cent of the parent intensity the minor peak sits on the shoulder of the isotope envelope regardless of resolving power.
— T. Oyelowo, Abeokuta
Yes, and this is the more useful statement of the problem. Resolving power is necessary and not sufficient; dynamic range and separation matter as much. We have added a sentence to the table note.
Lot-level verification with a public report is a real and achievable thing. It exists in this market, on a minority of listings.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Reported from the analysis, not from a warning notice.
We set out what is known, what is inferred and what is simply assumed about the fortnight after a vial is opened.
We looked at what the dose-ranging data supports about going higher, and it is thinner and less flattering than the market assumes.
The most useful single addition to any purity determination is a second separation under a different pH or on a different stationary phase, and the second-most useful is a…