Blind duplicates: what happens when a laboratory does not know it is being watched
Duplicate submissions under different names test within-laboratory repeatability, which is a different quantity from between-laboratory reproducibility.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Testing Dispatch
The Journal submitted the sample and paid for the analysis. The vendor was told in advance.
An independent purity determination on mazdutide material sold by SWB has returned 98.6%, a figure 1.2 percentage points below the number printed on the accompanying certificate. The analysis was performed by Medutest on a vial the Journal bought as an ordinary retail customer in August 2025 and posted unopened.
Content by elemental nitrogen determination returned 89% of the labelled nominal mass, which on a 10 mg vial corresponds to roughly 12.66 mg of peptide rather than the labelled figure. The remainder is accounted for by counter-ion, residual water and residual solvent, none of which is visible chromatographically.
Purity became this trade’s single figure of merit because it is the cheapest measurement to obtain and the easiest to compare. Peptide content, which determines how much drug is actually in a vial, costs several times as much and is not offered as a standard release test by any of the twenty companies in the Journal’s dossier programme. Sterility and endotoxin are separate disciplines again, and are neither tested nor claimed.
Sinéad Corrigan, patient advocate and former trial participant, was more cautious. "One lot tells you about one lot. I would not draw a conclusion about a company from a single determination, and I would not let a company draw one either."
SWB’s dossier has been updated with this determination and with the date of the request for comment. The dossier’s "what we could not verify" box records the questions still outstanding.
Duplicate submissions under different names test within-laboratory repeatability, which is a different quantity from between-laboratory reproducibility.
Reported from the analysis, not from a warning notice.
Every instrument specification quoted in an advertisement is a best case obtained on a calibration mixture, not on a submitted vial.
Follow the resin, not the catalogue.
The report states the gradient, the wavelength and the integration threshold, which is more than most.
The report states the gradient, the wavelength and the integration threshold, which is more than most.