Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Container closure

What VendorInvestigate is able to certify about a sealed vial, and what it is not

The Journal has read several hundred certificates from twenty companies. We set out what the documents actually cover, and what a reader is filling in from imagination.

A certificate of analysis for a research peptide is, in the overwhelming majority of cases, a document about chemistry. It reports a chromatographic purity figure, sometimes a mass, occasionally a water content, and it is generated from a few milligrams of powder dissolved in a vial of mobile phase. Nothing in that procedure has any bearing on whether the sealed container it came from holds viable microorganisms, whether the powder carries pyrogenic material from an upstream water system, or whether the stopper has maintained a seal since the day it was crimped. The document is accurate and it is answering a different question.

Two questions that share a piece of paper

Consider what happens physically when a certificate is produced. A few milligrams of lyophilised powder are weighed, dissolved in an aqueous mobile phase with an organic modifier, injected onto a reversed-phase column and separated over a programmed gradient while an ultraviolet detector records absorbance. The output is a trace. Software integrates the areas beneath its features, and the main peak area as a proportion of the total becomes the purity figure.

Every step of that procedure is blind to microbial contamination. A bacterial cell contributes no ultraviolet-absorbing peak at any retention time a peptide method would record. Endotoxin, a lipopolysaccharide, is not usefully detected at the wavelengths used for peptide bond absorbance and would in any case be present at a mass fraction several orders of magnitude below any integration threshold in commercial use. A vial holding a hundred colony-forming units and a vial holding none produce chromatograms that no analyst could distinguish.

This is not a defect of the method. Reversed-phase chromatography is an excellent way of determining what proportion of the chromatographically visible material is the intended species, and that is what it is being asked. The defect is in the reading. A document answering one question is being filed as evidence about five.

Bioburden, and why the number before the filter matters

A sterilising filter is not an unlimited barrier. Its qualification is expressed as a retention capability under a defined challenge — conventionally a high titre of a small bacterial species per square centimetre of membrane — and its performance in use depends on the load presented to it. A bulk solution carrying a heavy microbial burden presents a filter with a harder problem than one carrying a light burden, and it presents a second problem the filter cannot address at all: the endotoxin released by organisms that die upstream passes through the membrane unimpeded.

Regulated manufacture therefore specifies a pre-filtration bioburden limit, tests against it on every batch, and treats an excursion as an investigation rather than a curiosity. The specification is usually expressed in colony-forming units per hundred millilitres, and a well-controlled process runs far below it.

Of the twenty companies the Journal wrote to, four stated that pre-filtration bioburden is determined on every batch, three stated that it is determined periodically, and the remainder did not answer the question. We regard that distribution as the single most informative result of the correspondence, because bioburden testing is inexpensive, is performed on the bulk rather than the finished container, and is the earliest point at which a problem is cheap to fix.

A dye ingress test can pass a container with a defect large enough to admit an organism, because the dye happened not to travel.

On probabilistic leak testing

The filter, and the test that proves it survived

Sterilising-grade filtration through a membrane of 0.22 micron nominal rating is the operation on which aseptic processing rests. The membrane is qualified by challenge with a small bacterium at high concentration under the process conditions, and a filter that permits no passage under that challenge is accepted as sterilising for that fluid.

The critical practice is not the filtration but the integrity test that follows it. A membrane can be damaged during installation, during sterilisation, or by pressure excursions in use, and a damaged membrane looks exactly like an intact one. Bubble point, diffusive flow and pressure hold tests each detect a breach by measuring gas behaviour across a wetted membrane, and a post-use test is the only evidence that the filter was intact while the product was passing through it. A pre-use test alone establishes nothing about the state of the membrane at the end of the run.

This is the question the Journal has found most useful when assessing whether a fill operation is a real one. It is specific, it has a yes or no answer, the answer is recorded in the batch documentation as a matter of course, and it cannot be answered plausibly by anybody who is not actually running the process.

Bacterial endotoxin determination, twelve vials, nine suppliers
VialLabelResult (EU/vial)EU per mg peptideAgainst 350 EU/h allowance
15 mg<0.5<0.10Below quantitation
25 mg1.20.240.3%
310 mg2.80.280.8%
45 mg<0.5<0.10Below quantitation
55 mg14.62.924.2%
610 mg3.10.310.9%
715 mg6.40.431.8%
85 mg38.27.6410.9%
95 mg0.90.180.3%
1010 mg4.70.471.3%
115 mg<0.5<0.10Below quantitation
125 mg112.022.4032.0%
Single determinations by kinetic chromogenic assay at one accredited contract laboratory; method suitability established for each matrix; results expressed per vial as received and per mg of labelled nominal mass. The final column expresses the whole vial against the hourly allowance for a 70 kg subject and is arithmetic, not a safety assessment. Twelve vials from nine suppliers is not a survey.

Low endotoxin recovery, and the clean result that is wrong

The most uncomfortable finding in endotoxin testing over the past fifteen years is that certain formulation matrices cause added endotoxin to become undetectable over time. Spike a known quantity of reference endotoxin into a solution containing a non-ionic surfactant and a chelating buffer, hold it, and the recoverable endotoxin declines — sometimes to a small fraction of what was added — while nothing has been removed. The lipopolysaccharide aggregate state has changed, and the assay cannot see what it cannot bind.

The phenomenon is called low endotoxin recovery, and it matters because the combination of polysorbate with citrate or phosphate is extremely common in peptide and protein formulations. A hold-time study — spiking the actual product matrix and measuring recovery across the intended sample storage interval — is the standard mitigation, and it is now expected as part of method suitability for products in that formulation space.

The implication for a reader is narrow but worth stating. A negative endotoxin result on a surfactant-containing formulation, reported without a hold-time recovery study, is weaker evidence than it appears. A negative result on a reconstituted lyophilised peptide in plain water for injection, tested promptly, is considerably stronger.1

Depyrogenation, and the step that gets skipped

Because endotoxin survives sterilisation, glass components require a separate treatment. The standard is dry heat: a tunnel or oven cycle at two hundred and fifty degrees or above, validated to achieve at least a three-log reduction in a deliberately applied endotoxin challenge. Vials emerging from a qualified depyrogenation tunnel are both sterile and pyrogen-free, and they remain so only if the subsequent handling maintains it.

Elastomeric closures cannot take that treatment and are instead washed, siliconised and steam-sterilised by a validated cycle, with endotoxin control achieved by the washing step and specified as a limit per stopper. Water for injection is controlled at source, with a compendial endotoxin specification, and a water system is the commonest origin of a pyrogen problem in an otherwise competent facility.

Depyrogenation is also, in the Journal’s experience of asking, the step most frequently absent from descriptions of small fill operations. Autoclaved vials are sterile. Autoclaved vials are not depyrogenated, and an operation that describes its glass preparation solely in terms of autoclaving has told you something specific about what it has not done. We put this point to eleven correspondents; two said their glass is depyrogenated by dry heat with a validated cycle, and one asked us what depyrogenation was, which we took as a straight answer and a useful one.

2519136.300.110.2420.2830.142.9250.3160.4377.6480.1890.47100.11122.412EU per mg
Figure. Bacterial endotoxin per milligram of labelled peptide, twelve vials submitted by the Journal. Bars at the floor are results below the quantitation limit of the assay. Single determinations; not a survey.

What we submitted, and what came back

Between the second and fourth quarters the Journal purchased fourteen vials of lyophilised research peptide from nine suppliers, at catalogue prices, through ordinary channels and without identifying ourselves. Each was photographed sealed, logged, and stored at two to eight degrees on arrival. Twelve were submitted to a contract laboratory accredited to the general competence standard for testing laboratories for bacterial endotoxin determination by kinetic chromogenic assay, with method suitability established for each matrix. Two were submitted for a compendial sterility test by membrane filtration, which destroyed both.

We disclose the following limitations without being asked. Fourteen vials from nine suppliers is not a survey. Single determinations carry the uncertainty of single determinations. A negative sterility result on one vial says nothing about the batch it came from, for exactly the statistical reasons set out above. And a vial that has crossed a border in a padded envelope has a thermal and mechanical history we cannot reconstruct.

What the exercise establishes is narrower than a survey and, we think, worth publishing: that the tests exist, that they are commercially available to a private purchaser at a known price, that the numbers they return are interpretable against a calculable limit, and that nothing prevented any of the nine suppliers from commissioning them first.2

Documentation practice, named and criticised

The Journal tracks the release documentation of twenty companies. On sterility and endotoxin the picture is close to uniform. Every one publishes a purity figure. A minority publish an identity confirmation. Two publish peptide content. On the microbiological attributes, the standard document is silent, and the silence is not annotated: there is no line reading that sterility has not been determined, which would at least be informative.

Some of the practices we would like to see are already in use somewhere in the group. SSA reports peptide content on its certificates, having begun after correspondence with this publication. CPC and SWB describe their fill environment in general terms on request. QST and BCH answered our five questions in full. WXT and FGP declined on the grounds that research-use products are not represented as sterile injectables, which is a legally sound answer that concedes the point of the exercise. Several others did not respond, and we record non-response as non-response rather than as evasion.

What we are criticising is a documentary convention, not the conduct of any company named here. None of the twenty has been shown to us to have misrepresented anything. The convention is that a chemistry certificate stands in for a release package, and it is a convention this trade adopted collectively and could abandon the same way.

Twenty containers, fourteen days, destroyed. That is the entire empirical basis of the finished-product sterility claim.

On the compendial sterility test

Five lines that would change what a certificate is worth

Nothing the Journal asks for here requires a regulator, an inspectorate or a change in the law, and none of it is commercially sensitive. All five items are already known to whoever released the batch.

First, state the fill route: aseptic or terminal, and if aseptic, in what class of environment. Second, state the pre-filtration bioburden result against its specification, or state that bioburden is not determined. Third, state whether the sterilising filter was integrity-tested after use. Fourth, report bacterial endotoxin per milligram of peptide, with the method and the inhibition-enhancement result, or state that it was not determined. Fifth, where a sterility test has been performed, state the batch size, the number of containers tested and the method, so that a reader can do the arithmetic in the section above. Every one of the five is an ordinary element of a release specification for a biological product, and none of them is an invention of this publication.3

A certificate carrying those five lines would remain a one-page document and would be worth several times what the current one is worth, principally because four of the five lines are permitted to say no. A stated negative is a fact a reader can use. An omission is a space a reader fills with an assumption, and the assumption is always more favourable than the truth would have been.

A note on method and sourcing

The compendial material in this article is drawn from the current general chapters of the United States Pharmacopeia and the European Pharmacopoeia, read in the original rather than in summary, and from the international standards on aseptic processing and on laboratory competence. Where a chapter has changed status recently — as the recombinant reagent chapters have — we say so, because a reader consulting an older edition will find a different framing.

Where the Journal reports a number it obtained itself, it states the laboratory’s accreditation status, the method family, whether method suitability was established, and the number of determinations. Where we report what a company told us, we distinguish an answer from a refusal and a refusal from a non-response, because those three things are routinely collapsed in coverage of this trade and they are not the same.

Corrections to this department are handled by the standards desk, which reads every letter and records the outcome in the log. Readers who believe a paragraph here overstates its evidence are asked to write to standards@compoundjournal.com; readers with documents to send, including certificates they would like read, should write to letters@compoundjournal.com. We do not publish correspondents’ names without permission and we do not identify the source of a certificate.

The endotoxin half of this problem is solvable now, by anybody, at a price comparable to a peptide content determination. That is the single most practically useful sentence in this article. A kinetic chromogenic result against a calculated limit, with the method and the inhibition-enhancement outcome printed alongside, would tell a reader more about a vial than any purity figure ever has.

References

  1. “Low endotoxin recovery in biopharmaceutical formulations: mechanisms and hold-time study design.” PDA Journal of Pharmaceutical Science and Technology. 2017;71(6):452–467.
  2. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  3. International Council for Harmonisation. Q6B: Specifications — Test Procedures and Acceptance Criteria for Biotechnological/Biological Products. 1999.
  4. United States Pharmacopeia. General Chapter ⟨1207⟩ Package Integrity Evaluation — Sterile Products. USP–NF, Rockville, MD.
  5. United States Pharmacopeia. General Chapter ⟨86⟩ Bacterial Endotoxins Test Using Recombinant Reagents. USP–NF, Rockville, MD.
  6. “Cake appearance, collapse temperature and residual moisture in lyophilised peptide formulations.” AAPS PharmSciTech. 2019;20(7):286.

Related coverage

Analytics

What the nitrogen said

Peptide content is the number that tells you how much drug is in the vial. Almost nobody sells it, almost nobody buys it, and when we paid for it on nine vials the results…

Callum Brathwaite·17 Sep 2024·5 min