Where the solvent front ends is a decision, not a fact

We work through a single chromatogram twice, under two integration conventions, and show where the difference comes from.

The threshold is the parameter with the largest effect. Below it a feature is treated as noise and does not enter the calculation; above it, it becomes an impurity subtracting from the reported purity. A threshold at a tenth of one per cent of the main peak area discards a population of small related species that a threshold at two-hundredths of a per cent would report individually. Both settings are defensible for their purposes. Only one of them is capable of producing a purity figure below 99 on a well-made peptide, which is a fact about the parameter rather than about the powder.

The threshold, and what falls below it

Integration software applies a threshold — expressed as a slope sensitivity, an area cut-off, a height cut-off or some combination — below which a feature in the trace is treated as baseline noise and not integrated. The setting is necessary: without it, every fluctuation would be reported as an impurity and the result would be dominated by noise. The setting is also consequential, because a great many real, small, closely related species live in the region between the two conventional choices.

The arithmetic is easy to underestimate. A well-made peptide preparation may carry twenty or thirty related species each between two-hundredths and a tenth of one per cent — deletion sequences, deamidated and oxidised forms, epimers. Reported individually against a low threshold they might total a percentage point or more. Discarded against a high threshold they total zero. Two laboratories reporting 99.4 and 98.3 on the same lot may have measured the same chromatogram and disagreed only about which features are noise.

The Journal has asked all four independent services what threshold their standard peptide report uses. Two answered with a figure. One answered that it depends on the method and offered to supply the value per report, which is a better answer than a fixed number. One did not answer. We regard the threshold as second only to the gradient in importance and, like the gradient, it is a single value that whoever produced the document already knows.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.1

Retention-time agreement is consistency. Molecular mass is composition. Only fragmentation approaches sequence.

On three claims that share one phrase

What we submitted, and how we designed it

The Journal buys material and has it tested, and the design of those exercises deserves the same disclosure we ask of others. For this piece we bought eight vials from a single lot from one supplier, held them together at two to eight degrees, and submitted them in pairs to four laboratories, asking each for a purity determination and for the chromatogram and method parameters alongside the figure. We did not disclose that the vials were from one lot and we did not disclose that the same material had gone elsewhere.

Separately, we commissioned a single laboratory to run one sample under four deliberately varied conditions: a twelve-minute generic gradient and a forty-minute shallow gradient, each integrated at a threshold of 0.1 per cent and 0.05 per cent. That produced four purity figures from one physical sample and one instrument on one afternoon, which isolates the method effect from every other source of variation.

The limitations are ours to state. One lot from one supplier is not a survey of the market. Single injections carry the variability of single injections. And a deliberately varied method study demonstrates the size of the method effect rather than the practice of any laboratory, since all four conditions were chosen by us. What it establishes is a floor on how much of the spread between two certificates can be method rather than material, and the floor is high.2

Detection wavelength and what responds to it
WavelengthPrincipal absorberSees fragments without aromaticsBaseline noiseTypical use
214 nmAmide bondYesHigherPeptide purity and related substances
220 nmAmide bondYesModeratePeptide purity, quieter baseline
254 nmAromatic systemsNoLowSmall-molecule work, legacy detectors
280 nmTrp, Tyr, Phe side chainsNoLowTracking an aromatic residue; not a purity method
Diode array, 200–400 nmAll of the aboveYesMethod-dependentPeak purity assessment, spectral homogeneity
A purity figure generated at 280 nm is not comparable with one generated at 214 nm, and the difference is not a matter of a percentage point. Certificates reading at 280 nm circulate in this market; readers have sent us several.

Documentation practice, named

Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.

Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.

The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.

Dissolution, pH and what happens before the injection

A peptide in a vial exists in whatever state the manufacturer left it. Reconstituted in water, a hydrophobic sequence may not dissolve completely and the chromatogram will show particles or aggregates. Reconstituted in an acidic buffer, the same sequence dissolves and the chromatogram shows monomer. The purity figure—and the mass balance—changes accordingly. Sample preparation is not one of the twelve values that belong on a method disclosure, and yet it is one of the most consequential, because it determines what population the peptide is actually in when the injection happens.3

The trade addresses this by assuming that samples are dissolved in the mobile phase or its aqueous component, and the assumption is sometimes true and sometimes false. A certificate that states how the sample was dissolved, at what concentration, in what solvent and after what incubation time, is one that can be repeated. A certificate that does not is one that will produce different results if the receiving laboratory uses different dissolution practice, which is particularly consequential for poorly soluble sequences.

The technique is not on trial here and never was. Reversed-phase chromatography can resolve species differing by a single methyl group, and the laboratories running it for this market are, on the evidence we have gathered, largely competent and entirely willing to describe what they did when somebody asks. What is on trial is a document format that omits the four values needed to compare one figure with another.

References

  1. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  2. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  3. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.

Letters to the Editor

1 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

One practical note from the other side of the counter. When a customer asks us for the gradient we send it. When a customer asks a reseller, the reseller does not have it, because they were sent a PDF with a number on it. The gap you are describing is often two links down the chain rather than at the laboratory.

M. Halim, Kuala Lumpur

The Journal replies

That is an important structural point and it changes where the fix has to happen. If the laboratory report travels intact instead of being transcribed into a house certificate, the method travels with it. Four of the twenty companies we track already attach the original report, and on this argument they are doing the single most useful thing available.